4.1 Location

To select the installation site, follow the precautions described below:

■ Avoid places where fluid runs in a pulsating form.

■ Avoid places within the immediate proximity of equipment producing electrical

interference (such as motors, transformers, radio transmitters, electrolytic cells, or

other equipment causing electromagnetic or electrostatic interference).

■ Avoid places where excessive pipe vibration occurs.

■ Avoid places where there is direct sunlight. If this is unavoidable, use an appropriate

shade

■ Avoid places where corrosive atmospheres or high humidity conditions obtain.

■ Avoid places of too great an elevation or constricted areas where clearance

for installation or maintenance work is not provided.

■ Design piping so that the detector pipe is always filled with fluid, whether the fluid is

flowing or not.

■ The LF470 detector has no adjustable piping mechanism. Install an adjustable short

pipe where needed.

■ Chemical injections should be conducted on the downstream side of the flowmeter.

- 1 2 -