4.3 Piping Connections

(1) Required Pipe Length

If various joints are used upstream of the detector outlet, the straight pipe length as

shown in Table 4.2 is required.

Table 4.2 Required straight pipe length on the upstream side

L:

Required straight pipe length—straight pipe length plus half length of the detector.

D:

Nominal bore size (diameter)

NOTES

1.

The length of a reducer, if cormected, can be counted as a part of the straight pipe

length.

2.

No straight pipe length is needed on the downstream side. If a butterfly valve is

installed downstream of the detector, do not let the valve plate protrude into the

pipe of the detector

-