3. Use strong banding to secure scale in shipment.

2.3 SITE SELECTION

1. Extreme temperature variations may cause the scale to drift slightly off zero. Your precision
ton container scale has temperature compensated load cells to minimize such effects.
However, whenever possible, avoid placing your scale near air ducts, heating or cooling
elements, doors opening to the outside, ect. to minimize temperature effects.

2. Line power devices causing large inductive currents should not run off the same circuit as
the scale. Fluctuations in line voltage caused by such devices may result in display instability.

3. The hook-up cable to the read-out should not run close to other unshielded cables. Display
instability may result.

4. For best accuracy, a flat, level, and rigid surface is recommended to support the scale and ton
containers.

5. The area should be accessible for periodic cleaning.

2.4 INSTALLING

1. Clean the site area of dirt and debris.

2. Locate and mark all four corners. See Figure 1 for bolt/stud locations.

3

A 1/2" to 5/8" diameter bolt/stud is suggested. The bolt/stud should protrude 2" above ground.

4. Set scale into place.

52"

52"

FIG. 1

2.5 LEVELING

1. For best scale performance scale should be level within 1/2 degree.

2. Check all corners for wobble and shim if required.