1. Inaccurate but repeatable weight readings:

a.

Adjust span on read-out (see instrument manual)

2. Blank or drifting display:

a.

Consult the instrument manual.

b. Look for loose connection in hook-up cable at the instrument.

c.

Test for bad load cell (Section 4.1).

SECTION 4.0 SERVICING

4.1 Checking the load cell.

1. Load Cell resistance test:

a.

Disconnect the load cells from the instrument and measure the resistance as shown in

Figure 3.

b. Any electrical leakage between the leads and the load cell case is usually caused by moisture

leakage into the load cell or by moisture in a damaged load cell cable.

c.

If a load cell does not pass the above resistance tests, replace it with a known good one.

RED TO BLACK

360

GREEN TO WHITE

350 +/- 5

RED, BLACK, GREEN,

WHITE, OR YELLOW

GREATER THAN 200

TO CASE

MEGOHMS *

* Using a portable Ohm meter on highest range you should read infinity

LOAD CELL WIRING

BLK -EXT

GRN +SIG

RED +EXT

YEL SHLD

WHT -SIG

LOAD
CELL

LOAD CELL RESISTANCE CHECK

FIGURE 3

2. Load Cell zero shift test:

a.

Remove all the weight from the load cell and measure the output as shown in
Figure 4.

b. Connect a DC power supply of 10 or 15 volts to the Red (+) and Black (-) excitation load cell

leads.

c.

The measured output between the Green (+) and White (-) signal leads should be less
than 5 millivolts.