AUTO ZERO SCREEN

AUTO ZERO 120s

..

0.0mV ESC

This screen will automatically zero the cell input if there is a stable mV input for 120 seconds. If the
auto zero successful, the instrument will display an auto zero successful message. If fifteen minutes
pass without an auto zero, the instrument will display an auto zero failure message.

The top, right side of the display shows a timer, and the bottom, left side shows the current cell mV.

1

No function

2

No function

3

No function

4

ESC

Go back to the calibrate zero menu.

CALIBRATE CELL SPAN SCREEN

GET SAMPLE

0.0mV STR

ESC

1

No function

2

No function

3

STR

Sample the cell input mV & water temperature and enter the screen where the CL

2

residual value can be set.

4

ESC

Go back to the calibrate cell menu.

If one attempts to set a ppm when the mV input is lower than the zeroed mV, the error message
“ERROR! CELL mV <= ZERO” will be displayed for a few seconds.