The RA-1000 Residual Analyzer/Controller is an amperometric instrument
designed to continuously analyze residual levels of free or total chlorine, chlorine
dioxide, or potassium permanganate in water, wastewater or liquids containing
particulates.

II.

SPECIFICATIONS

Power Supply

115/230 VAC Switch selectable.

Power Consumption

12W

Display

LCD, backlit, 16 characters by 2 lines, 5mm character
height,5x8 character matrix.

Pushbuttons

Four, tactile dome.

Relays

High residual, low residual, and two spares
Mechanical (AC or DC), Rating (standard) 1.2A,
125VAC resistive or optional 5A @ 250VAC

Cell

Au/Cu (Gold anode/Copper Cathode) Galvanic Cell
input

Temperature

Thermistor 10kohms @ 25C, Beta=4100

Analog Output

Isolated, 4 to 20mA, 500ohm drive. Represents CL

2

residual

Digital Output

20mA Serial (transmits to remote computer)

Instrument Range:

0-0.1, 0-0.2, 0-0.3, 0-0.5, 0-1, 0-2, 0-3, 0-5, 0-10, 0-20,
0-30 mg/l field selectable

Sample Flow:

500 cc/minute: 5 psig maximum

Sample Supply:

Continuous. Where sample interruption may be required;
provisions must be made to keep the electrodes wet with
fresh water.

Analyzer Location:

As close as possible to the sample point to reduce lag
time.

Speed of Response:

4 Seconds from sample entry to display indication. Full
scale residual change 1-1/2 to 2 minutes.