Service Connection

Make sure that a 20 amp, 115 V outlet with the proper receptacle is available for the
10448 - EBC 47, 4-Ft Type A/B3. Both the 10276, 6-Ft Type A and the 10451, 6-Ft
Type B3 units require a 30 amp, 115 V outlet with the proper receptacle.

Connect the air, gas or vacuum service line to the external connection point located
on the top of the cabinet. The connections should be field tested for leak tightness.

Warnings and Other Identification Label Locations

Service Connection
Warning

Biohazard Warning

Service Card and Holder

Switch and Data Plate

Circuit Diagram (inside)

NSF Seal

Certification Warning

Flammable Materials Warning

FRONT VIEW

Figure 4: Top and Front Cabinet Views with Label Locations