5. Inspection and Maintenance

– 11 –

5.2 Cleaning

5.2.1 CFRP (Carbon Fiber Reinforced Plastic) Plate of the Sensor Unit

Disinfect the CFRP plate of the sensor unit each time a different patient uses the instru-
ment. If the optional grid unit is being used, disinfect its surface.

5.2.2 Cover

Clean the cover by the following procedure if it is dirty.

(1)

Turn OFF the power of the power box.
Shut down the control computer, and press side

“0”

of the power switch.

(2)

Turn OFF the power of each instrument.
Turn OFF the power of each instrument if connected.

(3)

Unplug the power cables.
Unplug the power cables of each instrument from the AC outlet.

(4)

Wipe the cover using neutral detergent.
Wipe the cover with a piece of cloth soaked in neutral detergent diluted in water and
wrung dry.

(5)

Wipe out neutral detergent.
Wipe the cover with a piece of cloth soaked in water and wrung dry whenever neur-
tral detergent has been used.

WARNING

Wipe the CFRP plate of the sensor unit with ethanol or glutaralde-
hyde solution to disinfect it each time a different patient uses the
instrument, in order to prevent infection.
If the optional grid unit is being used, disinfect its surface.
Please consult a specialist for the procedure for disinfection.
If you are using disinfectant other than those specified above, or
you are mixing another disinfectant with ethanol, please also con-
sult a specialist, because they may harm the CFRP plate and grid
unit.

WARNING

When the instrument is going to be cleaned, be sure to turn OFF
the power of each instrument, and unplug the power cable from
the AC outlet.
Never use alcohol, benzine, thinner or any other flammable clean-
ing agents. Otherwise, fire or electric shock may result.

!

!