9 Instant help

In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.


Help:

Fault

Possible cause

The displayed weight does not glow.

The balance is not switched on.

Batteries are inserted incorrectly or empty

No batteries inserted.

The displayed weight is permanently
changing

Draught/air movement

Table/floor vibrations

The weighing plate is in contact with

foreign matter.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing value is obviously

The display of the balance is not at zero

wrong

Adjustment is no longer correct.

Great fluctuations in temperature.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.