ABS-N_ABJ-NM-BA -e-1312

Confirm with TARE.

4. Return to weighing mode

Press ON/OFF repeatedly or for 3 sec.

5. Start tolerance control

If necessary, place an empty container on the balance and tare it by pressing
TARE

.

Place weighed goods and wait until the indicator

,

or

appears. With

the help of the indicator check if the weighed goods are under, inside or over the

default tolerance

Example of entry:

TG.VAL

100.0000 g

LM.VAL

10.0000 g

Weight less than nominal

weight and below the lower

tolerance

flashes slowly

flashes fast

Weight within tolerance [target

value ± tolerance]

(90.0000g – 110.000g)

Weight more than the nominal

weight and above the upper

tolerance

flashes fast

flashes slowly