4. Data format 4 „DF.4“
Example 9.9949 g:

Position

Data

ASCII code Explanation

1

S

53H

Stable weighing value = S (stable) 53H
Unstable weighing value = U (unstable) 55H
Positive weighing value = plus sign 2BH
Negative weighing value = minus sign 2DH

2

20H

3

20H

4

2BH

5

20H

Numeric weighing value is displayed in 8

positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2

positions with O L

6

+

2BH

7

9

39H

8

.

2EH

9

9

39H

10

9

39H

11

4

34H

12

9

39H

13

20H

Weighing unit

14

20H

15

g

67H

16

C/R

0DH

Separator C/R = 0DH, L/F = 0AH

At CR+LF data length will be increased


Printout examples KERN YKB-01N

S +9.9949 g Stable/positive weighing value

U +9.9949 g Instable/positive weighing value

S -9.9949 g Stable/negative weighing value

U -9.9949 g Instable/negative weighing value