1 adjustment with internal weight (kern abj) – KERN & SOHN ABJ 320-4NM User Manual

Page 24

Advertising
background image

24

ABS-N_ABJ-NM-BA -e-1312

8.2.1 Adjustment with internal weight (KERN ABJ)

With the internal adjustment weight, the weighing accuracy can be checked and re-

adjusted at any time.

Condition: Menu setting „I.CAL“.

When an optional printer is connected and the GLP function activated,

while starting the adjustment log „WAIT“ will be displayed. After the

printout the adjustment will be continued automatically.

Adjustment may be interrupted with ON/OFF, „ABORT“ is displayed.

Press the CAL-button and adjustment will take place

automatically.


The indicator

will be shown.


After successful adjustment the balance automatically

returns to weighing mode.

In case of an adjustment error (e.g. objects on the

weighing plate) the display will show an error message,

repeat adjustment.

When an optional printer is connected and the GLP

function is connected, the adjustment log will be edited,

see chpt. 8.3.

Advertising