12 Zero / tare menu (zeroing and taring functions)

Selectable functions:


1. Zero tracking function

+

see chpt 12.1

This function is used to correct automatically small

weight variations which appear directly after

switching-on.

In the event that small quantities are removed or added

to the material to be weighed, incorrect weighing results

can be displayed due to the “stability compensation“ in

the balance. (e.g. slow flow of liquids from a container

placed on the balance, evaporating processes).

When apportioning involves small variations of weight, it

is advisable to switch off this function.

2. Auto zero function

+

see chpt 12.2

This function is used to correct automatically small

weight variations which appear after a measurement

(e.g. soiled weighing plate) after the stability display.

3. Auto tare function

+

see chpt 12.3

After data output an automatic taring is carried out

4. Zero / tare timing

change function

+

see chpt 12.4

Selectable if the balance tares or gets set to zero

before or after indication of the stability mark.