13 Settings for Stability and Response

Exists the possibility to tune the stability of the display and the degree of reaction of

the balance to the requirements of certain applications or the environmental

conditions.

Most measurements can be carried out using default settings, that is, standard

mode. In standard weighing mode, stability and reaction have the same priority. For

certain applications such as e.g. dosage do use the dosing mode (pouring mode). In

dosing mode the reaction degree has the higher priority.

Beside the selection standard / dosing mode the stability of the display and the

reaction degree of the balance can additionally adapted in the menu.

Please note that data processing for greater stability generally slows the response

and processing for higher response reduces stability.

13.1 Stability and reaction settings via „Easy Setting display“

(without invoking menu)

In weighing mode press MENU shortly. The easy setting

display [

] flashes.

When the display is flashing, set stability or reaction via

UNIT

or PRINT as described in the following.

Priority on reaction

Priority on stability

Easy setting display

Operation

Every time the button is hit,

reaction will increase

Every time the button is hit,

stability will increase

The easy setting display flashes for a short time. Inputs are only possible

within this time. Using ON/OFF the flashing Easy setting display can be

switched off.