14.3 Formula mode
The formula function allows to add on various components of a mixture. For check

purpose, the weight of all components (CMP001, (CMP002 etc.), as well as the total

weight (TOTAL) can be printed out. The balance works with a separated memory for

the weight of the weighing container and of the formula components.

In formula mode the „Auto zero“ function is not active (see chpt. 12.2).


1. Connect printer (see chpt. 15 „Data output“)

2. Call up menu

In weighing mode press the MENU button twice

3. Select formula mode

Press the navigation buttons ( ) repeatedly until

APL.FUNC“ is displayed.

Confirm with PRINT

Press the navigation buttons ( ) repeatedly

until „FORMULA“ is displayed. The formula symbol

[ ] appears.

Press TARE. The indicator

is displayed, from

here on the balance is in formula mode.

Enable the edition „component no.“(see chpt. 14.3.1)

and „Total“ (see chpt. 14.3.2).

4. Weighing components

If necessary, place an empty container on the balance

and tare it by pressing TARE.

Press PRINT , the symbol

goes out.

When the GLP function (see chpt. 8.3) is activated, the

headline will be printed.

Weigh first component.

Press PRINT.

When standstill control ( ) is completed, the weighing

value of the first component (CMP001) is issued to the

optional printer. The displayed value is added into the

total adding memory. This is followed by the automatic

taring, the symbol [

] is displayed.