82

ABS-N_ABJ-NM-BA -e-1312

16.3 Communication parameters
By calling-up a standard setting „MODE 1 - MODE 5“

all communication parameters

are already preset (see chpt. 16.3.1).

The subsequent standard setting must be selected according to the printer (details

see the following table).

In the menu item „MODE U“ all parameters can be set defined by user (see chpt.

16.3.2).

Standard

setting 1

Standard

setting 2

Standard

setting 3

Standard

setting 4

Standard

setting 5

User-defined
settings

Setting for

KERN –

YKB-01N

Menu

selection

MODE 1

MODE 2

MODE 3

MODE 4

MODE 5

MODE U

MODE U

Manufactu

rer

Shimadzu

(Standard)

Shimadzu

*

Mettler

Sartorius

A-D

-

-

Baud Rate 1200

1200

2400

1200

2400

user-

defined

1200

Parity

None (8)

None (8)

Even (7)

Odd (7)

Even (7)

user-

defined

None (8)

Stop bit

1

1

2

2

2

user-

defined

1

Hand-

shake

Hardware Hardware off

Hardware off

user-

defined

off

Data

format

Shimadzu

Standard

Shimadzu

Standard

Mettler

Standard

Sartorius

Standard

A-D

Standard

user-

defined

DF.1

Separator C/R

C/R

C/R + L/F C/R + L/F C/R + L/F user-

defined

C/R

*only if the balance can send a feedback to the PC (without error: OK [C/R], at error NG [C/R].