16.4 Data Formats

In the menu item „D.FORM“ four data formats „DF.1-DF.4“ are available.

+

Menu setting, see chpt. 16.3.2, step 7 „Setting data format“.

1. Data format 1 „DF.1“
Example 9.9949 g:

Position

data

ASCII code Explanation

1

20H

Positive weighing value = space character 20H
Negative weighing value = minus sign 2DH

2

20H

Numeric weighing value is displayed in 8

positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2

positions with O L

3

20H

4

9

39H

5

.

2EH

6

9

39H

7

9

39H

8

4

34H

9

9

39H

10

g

67H

Weighing unit

11

20H

12

C/R

0DH

Separator C/R = 0DH, L/F = 0AH

At CR+LF data length will be increased


Printout examples KERN YKB-01N

9.9949g

Stable or instable/positive weighing value

-9.9949g

Stable or instable/negative weighing value