16.4 Data Formats
In the menu item „D.FORM“ four data formats „DF.1-DF.4“ are available.
+
Menu setting, see chpt. 16.3.2, step 7 „Setting data format“.
1. Data format 1 „DF.1“
Example 9.9949 g:
Position
data
ASCII code Explanation
1
20H
Positive weighing value = space character 20H
Negative weighing value = minus sign 2DH
2
20H
Numeric weighing value is displayed in 8
positions.
Not required positions = space character 20H
Where necessary, Overload is displayed in 2
positions with O L
3
20H
4
9
39H
5
.
2EH
6
9
39H
7
9
39H
8
4
34H
9
9
39H
10
g
67H
Weighing unit
11
20H
12
C/R
0DH
Separator C/R = 0DH, L/F = 0AH
At CR+LF data length will be increased
Printout examples KERN YKB-01N
9.9949g
Stable or instable/positive weighing value
-9.9949g
Stable or instable/negative weighing value