17.2 Service, maintenance
The appliance may only be opened by trained service technicians who are

authorized by KERN. Before opening, disconnect from power supply.

17.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to

valid national or regional law of the location where the appliance is used.

18 Instant help

Possible causes of errors:

In case of an error in the program process, briefly turn off the balance and

disconnect from power supply. The weighing process must then be restarted from

the beginning.

Fault

Possible cause

The displayed weight does

not appear

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Power supply interrupted.

The displayed weight is

permanently changing

Draught/air movement

Glass doors not closed

Table/floor vibrations

Weighing plate has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The weighing result is

obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

The balance is on an uneven surface.

Great fluctuations in temperature.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The desired weighing unit

cannot be called by UNIT

key.

Unit was not activated beforehand.

Automatic adjustment

carried out frequently.

Severe temperature variations in the room or the

instrument

No data transfer between

printer and balance.

Communication settings are wrong.

The menu settings cannot

be changed.

The menu is locked Remove the menu lock.