1 error messages – KERN & SOHN ABJ 320-4NM User Manual

Page 96

Advertising
background image

96

ABS-N_ABJ-NM-BA -e-1312

18.1 Error messages


Error message Explication

Remedy

ERR H

Hardware error

Switch balance off and on

again. If the error message

remains inform manufacturer.

ERR C

High zero point shift during

adjustment

Use ON/OFF to return into

weighing mode. Restart

adjustment process.

Objects present on weighing

plate.

Missing weighing plate

CAL D

Display instable

Check ambient conditions

(draught, vibrations etc.) Use
ON/OFF

to return into weighing

mode. Restart adjustment

process.

ERR N

Error at numerical input (e. g.

wrong password)

Correct input

ERR W

Wrong application

The balance changes to the

previous state. Correct

application.

COM ERR

Wrong remote control order.

The balance changes to the

previous state. Correct remote

control order.

- OL

Missing weighing plate

Install weighing plate correctly

OL

Overload

Reduce load

ABORT

Process cancelled

WAIT

Wait for process

BUSY

When adjustment was started

objects were present on

weighing plate

Remove object and continue

adjustment

Should other error messages occur, switch balance off and then on again. If the error

message remains inform manufacturer.