KERN & SOHN FCB 24K2 User Manual

Page 34

Advertising
background image

rE Cr:

The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.

Take into account that the following remote control commands must be sent
without a subsequent CR LF.


s

Function: Stable weighing value for the weight is sent via the RS232

interface

w

Function: Weighing value for the weight (stable or unstable) is sent

via the RS232 interface

t

Function: No data are sent, the balance carries out the tare function.


h. Format for stable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
B

*

B B B B B B B B 0 . 0 B g B B CR LF


i. Format in case of fault
1 2 3 4 5 6 7 8 9 10 11 12 13 14
B B B B B B B E r r o

r CR

LF


j. Format for unstable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10

11 12 13 14 15

B

*

B B B B B B B B 0 . 0 B CR

LF



Symbols

B

*

Blank or M

B / 0 /. Blank / weight values / decimal dot, depending on weighing value

g

Weight unit / pcs / %

E, o, r ASCII code or “E, o, r”

CR

Carriage return symbol

LF

Line feed symbol


FCB-BA-e-0911

34