7.1.2 Adjusting verifiable models

The adjustment is locked for verified balances.

with adjustment switch

To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chapter 7.2.1

Attention:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.


Display

Operation

 Start balance by pressing

While the balance carries out a self test (00…- 99…), press

and

keep pressed until „F1 CAL“ appears on the

display.

 Actuate the adjustment switch on the lower side of the

balance

 Press

in the display appears „UnLod“.

(Beispiel)

 Press

anew

The weighing value is displayed.

 Press

to enter the value of the adjustment weight

(see chapter 1)

 Place the adjustment weight

 Press

While the balance carries out a self-test, take away the
adjustment weight

The balance changes over to zero display.
Now the process of adjustment is completed.