Instant help – KERN & SOHN FOB 5K1S User Manual

Page 17

Advertising
background image

17

FOB-S-BA-e-1311

11 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.


Help:

Fault

Possible cause

The displayed weight
does not glow.

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Power supply interrupted.

(Rechargeable) batteries are inserted incorrectly or

empty

No (rechargeable) batteries inserted.

The displayed weight is
permanently changing

Draught/air movement

Table/floor vibrations

Weighing pan has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The weighing result is
obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.



Advertising
This manual is related to the following products: