KERN & SOHN FOB 5K1S User Manual

Page 9

Advertising
background image

9

FOB-S-BA-e-1311

Procedure when adjusting:

Observe stable environmental conditions.

A warming up time (see chapter 1; Technical Data) is required for stabilization.

Ensure that there are no objects on the weighing plate.


Balance display

Operation

 Start balance by pressing

The balance carries out an autotest and changes
then into the weighing mode, the zero display „0.0“
appears







(example)




(example)



 Press

for approx. 3 seconds, „CAL“ appears

shortly, then „CAL 0“ and after that „LoAd“ is
displayed alternately with the value of the first
adjustment weight.

 Place the first adjustment weight in the middle of the

weighing plate.


After a short time the value of the second
adjustment weight is displayed alternately with
„LoAd“

 Remove the the first adjustment weight from the

weighing plate and place the second adjustment
weight in the middle of the weighing plate.
Adjustment occurs automatically after standstill
control.

„donE“ will be shortly displayed. Then the value of
the second adjustment weight will be displayed.


Adjustment process has been finished successfully.
The balance is now in weighing mode.