RFB-BA-e-1211

18

6.8 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.

Verified models:

The adjustment is locked for verified balances.

To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chap. 6.9.1

Attention:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.


Operation

Display

 Switch on balance
 While the balance carries out

a self test, press

.