11 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.

Fault

Possible cause

The displayed weight does not glow.

The balance is not switched on.

The mains supply connection has been

interrupted (mains cable not plugged
in/faulty).

Power supply interrupted.

Batteries are inserted incorrectly or empty

No batteries inserted.

The displayed weight is permanently
changing

Draught/air movement

Table/floor vibrations

Weighing plate has contact with other

objects.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing result is obviously
incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.