10.2 Menu overview

F0

Not documented

,

settings are not allowed to be modified

Press TARE key

F1

Not documented

,

settings are not allowed to be modified

Press TARE key

F2

Not documented

,

settings are not allowed to be modified

Press TARE key

F3

Adjustment (see chap. 10.3)

Press TARE key

F4

Not documented

,

settings are not allowed to be modified

Press TARE key

F5

Not documented

,

settings are not allowed to be modified

Press TARE key

F6

Not documented, settings are not allowed to be modified

Press TARE key

F7

Not documented

,

settings are not allowed to be modified

Press TARE key

F8

Not documented

,

settings are not allowed to be modified

10.3 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature.
To receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.