Procedure when adjusting:
Attention:

In calibrated balances the adjustment function is switch locked.
To carry out calibrations, you will have to throw the unlock switch (See chpt. 11.1).
Observe stable environmental conditions. A warming up time of ca. 30 minutes is
required for stabilization. Ensure that there are no objects on the weighing plate.

• Activate the calibrating function F3 in the menu (See chpt. 10)

Weight

Basic price

Sales price

F3 WIDE

CAL.

• Press -key, the following display will appear:

Weight

Basic price

Sales price

0,000 kg

XXXXXX

(internal value)

• Ensure that there are no objects on the weighing plate.

• Operate menu activation key (10.1); stop check is carried out, then the weight

value of the calibrating weight will appear on the weight display

Weight

Basic price

Sales price

3.000

(maximum load)

Kg

(weighing unit)

XXXXXX

(internal value)

• Carefully place adjusting weight in the centre of the weighing plate

• Operate menu activation key (see chap. 10.1)

Weight

Basic price

Sales price

CAL.

done

• While the balance is counting down to zero, remove the adjusting weight.

• The balance returns automatically into weighing mode. An error message will

appear on the display should a adjustment error occur or should the adjust-
ment weight be incorrect. Turn balance off, then restart it and repeat the ad-
justment process.



Notice: Again throw the verification switch for calibrated balances.