3 adjustment of models cxb_m – KERN & SOHN CXP Version 1.7 User Manual

Page 23

Advertising
background image

7.6.3 Adjustment of models CXB_M


Operation

Ensure that there are no objects on the weighing plate.

 Press

+ 1 + 5 keys

 Press

key

The display shows:

0.000

M+

0

12345

 Press

key

The display shows:

15.000

(Example)

M+

0

12345

Weight value of adjustment weight is flashing on display.

 Carefully place adjusting weight in the centre of the weighing plate.

 Press

key

Adjustment will be carried out automatically after dead stop control.

 Remove adjustment weight during count down to zero.

The balance returns automatically into weighing mode. An error message will ap-
pear on the display should a adjustment error occur or should the adjustment
weight be incorrect. Turn balance off, then restart it and repeat the adjustment
process.


* The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Weights of different nominal values may be used for ad-
justment but are not optimal for technical measuring.
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com

CXB/CXP-BA-e-1417

23