KERN & SOHN EOA 100K-2 User Manual

Page 13

Advertising
background image

EOA-BA-e-1310

13

7.4 Plus/minus weighing


Operation

Display

 Turn on weighing scale

and wait for zero to

be displayed.

 Put the nominal weight on the weighing plate and tare

to [0] using the

button.

(example)

 Remove the nominal weight.

The nominal weight appears as negative display.

(example)

 Put the test objects subsequently on the weighing

plate, the respective deviation from the nominal weight
is displayed with the respective sign to „+“ and „-“.

(example)

 Back to weighing mode by pressing the

button and with unloaded weighing plate