9 Service documentation

• This chapter is only intended for a balance specialist!

• The weighing bridges are carried out in DMS sensor technology, at

every corner a DMS weighing cell is installed.

• The analogue-digital transformation occurs in the display unit. Also all

the balance and country-specific data are stored there.


9.1 Overview, setting regulation, tolerances

Testing and setting regulations:

Capacity

1500 kg

Readability

500 g

Min

10 kg

Max

1500 kg

1/3 corner load

500 kg

Tolerance

500 g

Verification data and tolerances as per OIML



-1

-0,75

-0,5

-0,25

0

0,25

0,5

0,75

1

0

500

1000

1500

[kg]

[g]

1500 kg

13

KFP V40-IA-e-1412