10 preload, deadload and overload settings – KERN & SOHN BBN 3T-3M User Manual

Page 16

Advertising
background image

10 Preload, Deadload and Overload settings

Kern model

Deadload** (kg)

**= already applied preload

Center Overload
Protection circa (kg)

Corner Overload
Protection circa (kg)

Loadcell
Capacity (kg)

KFP 600V40SM

90

NA

N/A

500kg

KFP 1500V40SM

90

NA

N/A

1000kg

KFP 1500V40M

120

NA

N/A

1000kg

KFP 3000V40M

120

NA

N/A

1500kg



Platform type

Platform dimension
(mm)

Loadcell

TC

Class

Max

E

max

E

min

Y

n

Dead-load

T

min

T

max

Cable-

Type

No.

Preload -1

-4

-3

(kg)

-5

-6

length

(kg)

(kg)

(g)

(m)

KFP
600V40SM

1000x1000x80

CZL-8C-
0.5T

D09-03.19

C3

300

500

0

10000

3000

300

-10 40

5

KFP
1500V40SM

1000x1000x80

SQB

TC6911

C3

750

1000

0

10000

3000

750

-10

40

5

KFP
1500V40M

1500x1250x90

SQB

TC6911

C3

750

1000

0

10000

3000

750

-10

40

5

KFP
3000V40M

1500x1250x80

SQB

TC6911

C3

1500

1500

0

10000

3000

1500

-10

40

5


16

KFP V40-IA-e-1412

Advertising