KERN & SOHN NFB 1.5T0.5LM Platform User Manual

Page 13

Advertising
background image

13

KFD V20-IA-e-1312

9.5 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.


Help:

Fault

Possible cause

The displayed weight is permanently
changing

Draught/air movement

Floor vibrations

Weighing plate has contact with other

objects.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing result is obviously
incorrect

No zero display with unloaded balance

Adjustment is no longer correct.

Great fluctuations in temperature.

Weighing bridge on an uneven surface.

Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.