14

10 Service documentation

This chapter is only intended for a balance specialist!

At every corner of the weighing bridge a DMS weighing cell is installed.

The analogue-digital transformation occurs in the display unit. Also all
the balance and country-specific data are stored there.


10.1 Overview, setting regulation, tolerances

Testing and setting regulations:

Capacity

600 kg

1500 kg

Readability

200 g

500 g

Min

4 kg

10 kg

Max

600 kg

1500 kg

1/3 corner load

200 kg

500 kg

Tolerance

200 g

500 g