9.2 Check and adjustment of the corner load

Check of the corner load:

Place a suitable aid such as a pallet on both weighing

beams. Make sure that the aid is designed to bear

the weight of the test weight.

Place the test weights in the centre of the load plate

and tare.


The balance displays -0-.

Place the test weights successively on all four

corners.

Now the deviations are displayed with sign, write

down the values. If there are deviations out of the

tolerances (see chap. 9.1), an adjustment will be

necessary.

Adjustment of corner load:

Preparation:

For a better control of the modifications which occur

during adjustment, select in the configuration menu

the highest readability for control purposes.

Open connection box

Adjustment rule:

The corner (weighing cell) with the biggest negative

deviation must be set to zero. Do not re-adjust this

corner even after several adjustment sequences.

Adjustment on the analogue print

Adjustment of weighing cell J2 takes place at the potentiometer VR1.

Adjustment of weighing cell J3 takes place at the potentiometer VR2.

Adjustment of weighing cell J4 takes place at the potentiometer VR3.

Adjustment of weighing cell J5 takes place at the potentiometer VR4.

Increase the value turning to the right, reduce the value turning to the left.