6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use
The weighing beams are designed in a way that reliable weighing results are

achieved in common conditions of use.

To enable exact and fast work, select the right site.


On the installation site observe the following:

Place the weigh beam on a level and stable surface.

The foundations at the site must be able to bear the weight of the weighing

beams as well as the weight of the maximum weight.

Avoid extreme heat as well as temperature fluctuation caused by installing

next to a radiator or in the direct sunlight

Protect the weighing system against direct draughts due to open windows and

doors

Avoid jarring during weighing

Protect the weighing beams against high humidity, vapours and dust

Do not expose weighing beams to moisture over a long period of time. Non-

permitted condensation (condensation of air humidity on the appliance) may

occur if a cold appliance is taken to a considerably warmer environment. In

this case, acclimatize the disconnected appliance for ca. 2 hours at room

temperature

Avoid static charge of goods to be weighed or weighing container

Do not lean weighing beams against a wall

Do not move weighing beams whilst loaded

Keep away chemicals (such as liquids or gasses), which could attack and

damage the weighing beams inside or from outside

Keep IP protection of the device

Major display deviations (incorrect weighing results) may be experienced should

electromagnetic fields (e.g. due to mobile phones or radio equipment), static

electricity accumulations or instable power supply occur. Change location or remove

source of interference.