3 checking the original dimensions, 4 battery operation – KERN & SOHN CH 50K100 User Manual

Page 13

Advertising
background image

4.3

Checking the original dimensions

 Enter the original dimensions shown on the production data sheet in the grey

boxes of checklist chap. 8.3.

 Check original dimension of suspended balance; for implementation see chap. 8.3

„Regular Maintenance“

 Enter all data (date, tester, results) in the first line under „Inspection before first

use“ in the checklist (see chapter 8.3)

CAUTION

If the dimensions of your first safety inspection do not match those
of KERN, the balance must not be put into operation. In this case
please contact a service partner authorised by Messrs. KERN.

4.4

Battery operation

1 x 9 V block battery



Push out battery cover on the rear side of the balance in arrow direction. Insert V
block battery. Reinsert the battery cover. If the batteries are empty, in the balance
display appears “LO”. Press ON/OFF/TARE -key and replace the batteries
immediately. In order to save the battery, the balance switches automatically off after
4 minutes without weighing. This auto-off function can be deactivated in the menu
(see chap. 6).

CH-BA-e-1432