Error messages – KERN & SOHN HFB 600K200 User Manual

Page 26

Advertising
background image

HFB-BA-e-1212

26

8. Error messages


Error
message

Description

Possible causes

--ol-

Maximum load exceeded

 Reduce load
 Check whether the balance has

been damaged

Err 4

Zeroing range exceeded
(normally 4% max)

 Overload when zeroing

Err 5

Keyboard error

 Improper operation of the balance

Err 6

Value outside the A/D changer
range

 Weighing plate not installed
 Damaged weighing cell
 Damaged electronics

Err 9

At an instable weighing value
ZERO or TARE button
actuated

 Wait for a stable weighing value

Err 17

Tare value out of range

 Reduce tare load

Ba lo

Capacity of rechargeable
battery exhausted

 Recharge battery


Should other error messages occur, switch balance off and then on again. If the error