4.3

Checking the original dimensions

Enter the original dimensions shown on the production data sheet in the grey
boxes of checklist chap. 10.3.

Check original dimension of crane scales; for implementation see chap. 10.2
„Regular Maintenance“

Enter all data (date, tester, results) in the first line under „Inspection before first
use“ in the checklist (see chapter 10.3)

CAUTION

If the dimensions of your first safety inspection do not match those
of KERN, the balance must not be put into operation. In this case
please contact a service partner authorised by Messrs. KERN.

4.4

Rechargeable battery operation

CAUTION

Material damage on crane scales

Only use the delivered mains adapter.

Make sure that the mains adapter, the cable and the
mains plug are in a perfect condition.

Do not use the crane scales during the loading process.



Before the first use, the rechargeable battery should be charged by connecting it to
the mains power cable for at least 24 hours. The operating time of the rechargeable
battery is approx. 100 hours.
If the capacity of the rechargeable battery is exhausted, „LOb“ appears. The balance
will be ready to operate for about another 10 minute, then it will switch off
automatically. Connect the power cable as soon as possible to load the rechargeable
battery.

When the crane scales are out of operation for a longer period, remove the
rechargeable battery.