Procedure:

In weighing mode, press

repeatedly until [tCH] appears.

Press

, [Pin] is displayed

Press

,

and

subsequently, [P1 SPd] will appear

Press

, „P2 CAL“ will be displayed

Operate adjustment switch; for position see chap.15

Press

, [dESC] is displayed

Press

repeatedly

until [CAL] will be displayed.

Confirm with

, [UnloAd] appears

Ensure that there are no objects on the weighing pan.

Wait for stability display "STABLE", then confirm with

(example)

The size of the currently set adjustment weight will be
displayed.

To change, select the digit to be altered by

, and the

numerical value by

.

Confirm with

, [LoAd] appears.