5. Servicing

If there should appear some abnormal phenomena to the
instrument, please do not dismantle or adjust any fixed
assembly parts on your own. Instead of this, the present
warranty card has to be filled out and the instrument has to
be sent to us. The warranty service can be carried on.

6. Transport and Storage

1) The instrument has to be kept away from vibration,
strong magnetic fields, corrosive medium, dumpiness or
dust. Storage in ordinary temperature.

Appendix A Sound Velocities

Sound Velocity

Material

In/us

m/s

Aluminum

0.250

6340-6400

Steel, common

0.233

5920

Steel, stainless

0.226

5740

Brass

0.173

4399

Copper

0.186

4720

Iron

0.233

5930

Cast Iron

0.173-0.229

4400 5820

Lead

0.094

2400

Nylon

0.105

2680

Silver

0.142

3607

Gold

0.128

3251

Zinc

0.164

4170

Titanium

0.236

5990

Tin

0.117

2960

Epoxy resin

0.100

2540

Ice

0.157

3988

Nickel

0.222

5639

Plexiglass

0.106

2692

Polystyrene

0.092

2337

Porcelain

0.230

5842

PVC

0.094

2388

Quartz glass

0.222

5639

Rubber, vulcanized

0.091

2311

Teflon

0.056

1422

Water

0.058

1473




Appendix B Application Notes
Measuring pipe and tubing

When a piece of pipe is measured to determine the
thickness of the pipe wall, the orientation of the transducer
is of importance. If the diameter of the pipe is larger than
approximately 4 inches, measurement should be
performed with the transducer orientated in the way that
the gap in the surface of the sensor is perpendicular (at
right angle) to the long axis of the pipe.
For smaller pipe diameters, two measurements should be
performed, one with the surface gap of the sensor
perpendicular, another with the gap parallel to the long
axis of the pipe. The smaller one of the displayed values
should be taken as the thickness of that point.


Measuring hot surfaces

The sound velocity through a substance is dependent on
its temperature. As materials heat up, the velocity of sound
through them decreases. In most applications with surface
temperatures of less than 100°C, no special procedures
must be observed. At temperatures above that point, the
change in sound velocity of the material being measured
starts having a noticeable effect upon ultrasonic
measurement. At such elevated temperatures it is
recommended to first performing a calibration on a sample
piece of known thickness, which is at or near the
temperature of the material being measured. This will allow
the instrument to correctly calculate the sound velocity
through the hot material.
When performing measurements on hot surfaces, it may
also be necessary to use a specially constructed high-
temperature transducer. These transducers are built of
materials which can withstand high temperatures.
It is also recommended that the sensor has to be left in
contact with the surface for a short time in order to acquire
a stable measurement. While the transducer is in contact
with the hot surface, it will be heated up and with thermal
expansion and other effects, the accuracy of measurement
may adversely be affected.

Measuring laminated materials
Laminated materials are unique because of their density
(and therefore sound velocity) may considerably vary from
one piece to another. Some laminated materials may even
exhibit noticeable changes in sound velocity across a
single surface. The only way to a reliable measurement is
to perform a calibration on a sample piece of known
thickness. Ideally, this sample material should be a part of
the same piece being measured, or at least from the same
lamination batch. The effects of variation of sound velocity