PEAK:


Three functions available:
-

Track mode (continuous measurement)

-

Peak mode (capture of maximum values)

-

Auto-Peak mode, same as Peak-mode, only
without the „Min limit peak save“ function

MEMORY:

Saves the peak values to calculate the average
value (please see section

“Memory mode”



DELETE Function


Deletes stored peak values (only in

„Memory

mode

“ active)



PRINT:

Sends the stored peak values to a PC or Printer
(plea

se see section „Data Output”)



High / Low limit function


LEDs to display OK / NOT OK Tests

Lower than lower limit

Lights if the STOP value is reached

Higher than highest limit


This function allows efficient

testing of OK / NOT OK’S

measurements of similar or identical testing objects.
A lower and an upper limit value can be defined. The
instrument compares the individual measuring results with
these limit values and shows the OK or NOT OK result by
green or red light diode and by sound.
To set these limit values, please see the SET Menu in
section „Operating keys“.

Measurement (Track Modus)

Display (1) shows the continuous force in a defined
direction (6)

To zero the display, press:



Peak-Hold Function (Peak Mode)


Please press:

Auto-Peak-Hold-Function (Auto-Peak Mode)


Please press:


Min. limit peak function to activate storage of
measurements


This

function allows to eliminate unwanted „Pre-Peak

values“ that are lower than the main peak value (Fp). The
“Min limit peak save” value (Fo) takes care, that these
“Pre-Peak values” are not saved.
The „Min limit peak save“ function is only possible in Peak-
Mode.
To set this Min limit value, please see the SET Menu in
section „Operating keys“.
Memory mode and average value (from up to 10 peak

values)

Saving peak values in the instrument

Activating the „AUTO PEAK Function“ by PEAK key


Deactivating the „Average Function“ by MEMORY key


 Now, all peak values are stored automatically in the
instrument

To browse through the stored values, please use the ▲

or ▼ keys. (The values will be shown in the upper

display segment)

 By pressing the MEMORY key, the average value of the
stored peak values can be shown (in the upper display
segment)

To delete every stored value, press the▼-key in the

AVERAGE-Mode