5.

Operation

a.

Display


Measuring Result
Measuring Units
Measuring Direction
Indication of PEAK Mode

b. Operating keys


ON / OFF:
to switch on, ON has to be pressed for about 1s

MAX:
Changes between Track and Peak mode (capture of
maximum values)

UNIT:
Press shortly to select desired unit:
N, kg, lb, ou

ZERO:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)

c. Display Return of 180°


When the instrument itself is turned so that the head faces
down, the display returns automatically.

d. Measurement (Track Modus)


- Display (1) shows the continuous force in a defined
direction (3)
- To zero the display, press ZERO key:

e. Peak-Hold Function (Peak Mode)


Please press: MAX key








6. Backside Fixing to a test stand

Fixing by 4 x M3 screws.

7. Adjusting procedure


- Observe stable environmental conditions. A short
warming up time of approx. 1 minute is recommended for
stabilization.
- Switch on the unit by pressing the Unit-key and keep it
pressed until in the display first appears shortly "CAL".
- After that there appears the fix size of the adjustment
weight. Attach the adjustment weight; a short time later
appears "F". Then the instrument returns automatically to
the measuring mode.

In case of an adjustment error or incorrect adjusting weight
the display will show "E"; repeat adjustment process.

8. CE Declaration of Conformity