LABEC AA20 User Manual

Page 3

Advertising
background image

 
 
 
Location 
Select a location free from draught and away from direct sunlight or other heat source. Proximity to a drain

   

 

 

 

 

 

 

 

 

   

 

 

 

     

 

for: 
 
1.

Waste from the manually operated drain. 

2.

The outlets servicing the pressure release valve and air vent 

3.

Adjustable steam release valve at the bottom right hand rear of cabinet. 

 
 
General 
The LABEC Automatic Autoclave is designed and constructed to meet the requirements of AS1210 of the

 

 

 

   

 

 

   

 

 

   

   

 

Pressure Vessel Code and has been approved by N.S.W. Workcover Authority, to operate at pressures up

 

 

 

 

 

 

 

 

 

 

   

   

 

 

to 215 K.P.A. Regulations require that on installation the autoclave be examined by a “Competent Person” or

 

 

 

 

 

 

 

 

 

   

     

 

   

licensed boiler inspector and this certificate of inspection be displayed near the autoclave and renewed and

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

inspected annually. For more details contact your local equivalent of Workcover NSW or contact Labec for

 

 

 

 

 

 

 

 

   

 

   

 

 

 

more information. 
 
Approval Numbers 

 

NSW 

QLD 

SA 

WA 

TAS 

AA9 

6766­P­87 

 

 

WA94 

 

AA13/AA16 

674­P­90 

BV18716 

PV86636/7 

WA9489 

TA4946 

AA18/AA20 

673­P­90 

BV18717 

PV86636/7 

WA9490 

TA4947 

 
Heating 
Heating of the autoclave is by means of stainless tubular element located in the base of the autoclave

   

 

   

 

   

 

 

 

   

 

   

 

 

vessel. 
 
Description of Controls 
See Appendix A Attached 
 
Temperature Control 
Labec autoclaves are fitted with solid state proportional action temperature controllers which operate from a

 

 

 

 

 

 

 

 

 

 

 

 

 

   

sensitive thermocouple or Rtd inserted in the working space of the chamber. The controller has been

 

 

 

 

   

 

 

   

 

 

 

 

 

 

calibrated at 121°C and before any adjustments to temperature settings are made allow the autoclave

 

 

 

 

 

 

   

 

 

 

 

 

 

 

temperature to stabilise for at least one half hour. For the digital controller fitted please read the enclosed

   

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

operating instructions when setting the controller temperature. The timer will automatically start the

 

 

 

 

 

 

 

 

 

 

 

 

 

countdown process once the temperature reaches the set value. (OP4 indicates the timer has

 

 

 

 

 

 

 

 

 

 

 

 

 

 

initiated). 
 
Important Safety Instructions 
 
Caution 
Please observe the following safety measures before using your LABEC equipment. 
1.

Ensure pressure inside the autoclave is reading zero KPA on the pressure gauge before opening the

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

lid. 

2.

These autoclaves are NOT FLAME PROOF and under no circumstances should inflammable,

 

 

 

 

 

 

 

 

 

 

 

 

combustible or explosive material be placed in the autoclave. 

3.

Low ignition temperature materials and those materials which give off inflammable or explosive

 

 

 

 

 

 

 

 

 

 

   

 

vapors should not be placed in the autoclave. 

4.

Avoid heating substances which give off corrosive vapor. 

5.

Users are advised of the dangers of heating combustible materials. The manufacturer can

 

 

 

 

 

 

 

 

 

 

 

 

 

recommend special types of elements which will prohibit the autoclaves temperature  

6.

Observe those rules pertaining to wiring and installation of electrical appliances as recommended by

 

 

 

   

 

 

   

 

   

   

2 Laboratory Equipment Pty Ltd Instruction Manual ­ Autoclave AA13,16,18,20                                 Ver 2.51 2013 