LABEC AA20 User Manual

Page 6

Advertising
background image

 
 
Maintenance 
 
Check the Safety Valve monthly by pulling the safety spring for 1 second to release steam whilst

 

 

 

 

 

 

 

 

 

 

   

   

 

 

 

under pressure. Wear Full Face Mask, Gloves and full length clothing and other PPE as required.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ensure no product is inside the Autoclave during testing. 
 
Clean drain and flush with boiling clean water regularly to prevent build up. 
 
Ensure Pressure Gauge reads 105 KPA at 121ºc when in use. If not switch off immediately and have

 

 

 

 

 

   

 

   

   

 

 

 

 

 

 

the unit serviced. 
 
Have the unit inspected by a registered boiler inspector annually. ( Ensure chamber is registered

 

 

 

 

   

 

 

 

   

 

   

 

with the appropriate body for pressure vessels) 
 
The cabinet is finished with stainless steel and to maintain appearance should be wiped over with a non

 

   

 

 

 

 

   

 

 

 

 

 

 

   

 

abrasive cleanser. The chamber is manufactured from stainless steel and may be cleaned with a solvent

 

 

 

   

 

 

 

 

 

 

 

 

   

 

and if grained stainless steel chamber a stainless steel scratch pad may be used. The pad should be

   

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

rubbed in the same direction as the grain pattern of the stainless steel. 
The silicone gasket should be cleaned with detergent only, ensuring it is dried completely after washing. To

 

 

 

 

 

 

 

 

 

     

 

 

 

 

 

prolong life it may be stored at cool temperatures in a container of water. Gasket lubricant is available to

 

   

 

 

   

 

     

   

 

 

   

   

provide prolonged life and better sealing on the gasket. 
The locking pins should be regularly greased with high temperature grease to reduce binding in the guides.  
 
WARNING: THIS AUTOCLAVE SHOULD NEVER BE CONNECTED TO MAINS WATER SUPPLY  
 
Troubleshooting 
 

SYMPTOM 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check autoclave is plugged in and

 

   

   

power switched on. 
 

2. Ensure mains power supply point is

 

 

 

 

   

functioning by using a test appliance

   

   

 

on power socket. 
 

3. Check internal RCD has not tripped. 

 

 

Failure to heat or maintain temperature 
 
(Indicator light is on) 

1. The temperature controller is on. 

 

2. Handle ‘A’ is pulled forward activating

     

 

 

the microswitch. 
 

3. The water level is above the low

 

 

 

 

 

 

water sensor and the low water

 

 

 

 

 

 

indicating lamp is off. 
 

 

The timer does not start 

1. Check the AL4 value is equal to set

 

 

 

   

   

value as this value activates the

 

 

 

 

 

 

timer. 
 

 

 

5 Laboratory Equipment Pty Ltd Instruction Manual ­ Autoclave AA13,16,18,20                                 Ver 2.51 2013 