with knife or screwdriver. And there is a rope

behind the plastic plug, once you pull the rope, the door cover

can be open, and then you can remove the centrifugal samples out from the machine.
6.1.4 You must put the rope back in the machine and put the plastic plug back to original position after you
open the door cover.

6.2 Information of trouble alarm

The table shows the alarm information of the machine and the reason as well as the way to solve; you can

rule it out according to the table. IF users still can't solve the problem or the alarm information is not included,
please contact maintainer from our company immediately.

Please turn off the electricity first when there is any problem.

And start it until the problem is solved.

Table 3: Data of trouble alarm

Symbols

Information of abnormal
phenomenon

Troubleshooting

E2

Over speed. It stops to
work because of detection
of over speed of rotor

● Problem of microcomputer control system,

please contact us.

● Problem of speed sensor. Please contact

us.

E3

Door cover protecting

● Re-close the door cover
● Please contact us if the door cover is

broken

E5

parameter setting is wrong

● Please re-set according to content of

parameter setting

E7

Fault of driving

● contact our company please

E8

Parameter setting error

●Follow the parameter settings to re-set in

part