Warning, Caution, Read and save these instructions – Oakworks DTPM300 User Manual

Page 5: Important safety instructions

Advertising
background image

Improper use of this device can cause injury. Be sure to read all operating instructions prior to use.
Duty Cycle: Max. 10% or 2 min. continuous use at full load then 18 min. without use.
Use caution when changing the table position to avoid collision with other moving or stationary items.
The Foot control must remain on the floor. Damage to the table will result if the foot control is placed under the col-

umn.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service

according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of accessories, transducers, and cables other than those specified by the manufacturer, may result in increased

Emissions or decreased Immunity of the table.
The table should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the

Table should be observed to verify normal operation in the configuration in which it will be used.
The Imaging Table is designed to be a standalone table used with radiographic equipment. This table must not be modi-

fied or incorporated into any other equipment.
Weight limit (patient and accessories): 450 lbs. / 205 kg.
A pinch hazard exists under the side rail on either side of the table. The tabletop slides in a head to toe direction on a

roller track system. Be sure to keep fingers and arms away from the table’s slide mechanism.
The table top slide mechanism must be locked at all times except when the top is being moved. Both locking knobs must

be securely tightened at all times - Travel can occur if the locking knobs are not secured by inadvertently pushing on the

end of the table or by tilting the table.
The table utilizes four locking casters to permit movement of the table within the imaging suite. Accidental movement of

the table may occur. Lock at least two casters prior to accomplishing imaging of the patient.
When lowering the table or using the Trendelenburg functions, make sure there is nothing underneath the table top that

can impede motion (like stools, cabinets, accessory parts, cleaners, etc.)
Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory patient.

The patient may lose balance and fall.
The table is not to be used as a gurney or to move patients from room to room.
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
After unpacking your table, inspect it thoroughly for damage. If you suspect a problem, do not use the table and call our

customer service department at 717.235.6807.

WARnInG

Grounding reliability can only be achieved when this equipment is connected to an equivalent 3 prong receptacle marked

“Hospital Only” or “Hospital Grade”. See the Grounding section of this manual.
Patient must be secured with a Safety Restraint Strap prior to using the table.
Electrical Shock Hazard. The power supply/control module is located on the base of the table. No user serviceable parts

are inside the control box. Refer servicing to qualified personnel. Unplug wall connector prior to contact with any cables

connected to the power supply.

2

IMPORTAnT SAfeTy InSTRUCTIOnS

CAUTIOn

ReAD AnD SAve TheSe InSTRUCTIOnS

IMPORTAnT SAfeTy InSTRUCTIOnS

Advertising