Cooper Instruments & Systems DCM 433 Current Bridge Amplifier User Manual

Page 3

Advertising
background image

CF 26

3

4/2001

Specifications

NOTE: Unless otherwise noted, specifications apply after one hour warm up at 25°C ambient. Temperature
coefficients apply between 0°C and 55°C ambient.

Isolation

Input to Output

700 VDC
60pF

Amplifier
Gain
Adjustment

Range

Input 4-20mA Output

5 mV to 50 mV

Linearity

±0.01% of Full Scale

Temperature Stability

50 PPM (0.005%/°C)

Input Noise

.1 Hz to 10 Hz

2µV P-P

Tare Adjustment Range

Bridge

Offset

-3 mV to +25 mV
(Equals 80% F.S. of 3 mV/V cell)

Input Resistance

Differential and Common Mode

1000 megaohm

Common Mode Rejection, DC

100 dB minimum

Common Mode Input

+5 Volts

Output

0 or 4 mA

Zero Adjust
Temperature

Coefficient

1 µA/°C

Current Output Span

4 to 20 mA or 0 to 20 mA

Compliance Voltage

-1 to +20 Volts

Current

-1 mA to +20 mA

Loop Resistance

0 to 1000 ohms

Frequency Response, -3dB

10 Hertz, 2 Pole Roll Off

Response

Time

Rise Time 10% to 90%

To 0.1% of Final Value


35 ms
100 ms

Bridge Supply
Voltage Adjustment Range

5 to 10 Volts

Temperature Coefficient

100 PPM Typical

Fixed

10.2V

±2.5%

Temperature Coefficient

60 PPM Typical

Load Current

0 to 120 mA

Regulation - Load and Line

0.02% maximum

Output Noise

120 Hz Bandwidth

1 mV RMS, Maximum

Power Input

115 VAC, ±10V 50/60 Hz @7 VA

Optional

220 or 230 VAC, ±10%

Line Isolation
Capacitance

60

pF

Dielectric Withstand

1100 Volts RMS

Environment

Ambient Operating

-25°C to +55°C

Storage

-25°C to +85°C

Weight

18 oz. (510 grams)

Size

3.75” L x 2.0” W x 2.87” H
(9.53 cm x 5.1 cm x 7.62 cm)

Advertising