background image

CF 26

5

4/2001

Transducer Excitation

The bridge excitation supply voltage is set by SWB-1 and potentiometer E. Set SWB-1 ON for a fixed 10 Volts. This
will provide the best temperature stability. The supply can be adjusted between 5 and 10 Volts by setting SWB-1
OFF and adjusting potentiometer E. The supply will deliver up to 120mA current at any voltage setting to power up
to four 350-ohm sensors.

SWB-1

Excitation Voltage

ON

Fixed 10 Volts

OFF

Adjustable



SWA-1 Tare

Range

OFF

-3 mV to +15 mV

ON

+15 mV to +25 mV

SWA-2 Zero

Current

OFF 4

mA

ON 0

mA

SWA-4 SWA-5 Span

Range

ON

ON

40 mV to 50 mV

OFF

ON

20 mV to 40 mV

ON

OFF

10 mV to 20 mV

OFF

OFF

5 mV to 10 mV