Bi-directional rs-232 error messages – Scientech S9000 User Manual

Page 19

Advertising
background image

Format B: Balance in normal weighing mode

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17

X X X X X X X _ _ _ Y Y Y _ _ Y Y

...CR

LF

Display shows 5.15 g and balance in high range

Message is _ _ _5.15_ _ _G_ _HIGH CR LF

Format B: Balance in a special weighing mode

Display shows 1250 PCS and balance in low range

Message is _ _1250_ _ _ _ _PCS_ _LOW CR LF

Note that, in these special weighing modes, the least significant digit of the numeric field is in the 6th (or 7th position if the

number is negative) and that the leftmost character of the annunciator is in the 12th position.

Format C:

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23

R E G : _ X X X _ _ X X X X X X X _ _ _ Y Y CR

LF

Display shows the contents of register 24 as 102.56, an additive constant, with units in grams

Message is

REG:_024_ _ _102.56_ _ _G CR LF

Display shows 12:03:35 as the time

Message is

REG:_101_12:03:35_ _ _TIME CR LF

BI-DIRECTIONAL RS-232 ERROR MESSAGES

There are three single character error messages that the balance issues when certain errors occur during bi-directional

communications. Each error character is followed by CR LF.

1.

! is issued when the input buffer is full. Normally this occurs because a CR is lost or the command string is too long.

The balance has a 37 character input buffer which is sufficiently long for all valid command strings. The ! is issued as

soon as the buffer is full and is repeated after each successive character until a CR is received. In response to the CR, a

? and CR/LF is sent and the balance input buffer is cleared.

2.

: is issued when a parity, overrun, or framing error is detected by the interface since the last CR. The : is sent after the

CR ending the command string which caused the error has been received.

3.

? is issued when a syntax error occurs or when an illegal character is received or when some other procedural error is

detected. The ? is sent after the CR ending the command string which caused the error has been received.

Power up errors are not detectable over the RS-232 I/O channel. The only way to detect such errors is to attempt to

communicate with the balance and have the ability to recognize that the communication is failing. Displayed error messages are

only available by polling the balance using the SEND command. The balance would respond by transmitting the displayed error

(i.e., Err 1, Err 2).