Level the balance, Install the weighing pan, Connect to an electrical outlet – Scientech HC Series User Manual

Page 4: Turn the balance on

Advertising
background image

Level the Balance:

The balance has one fixed foot and three adjustable feet which are used to level the balance by centering the bubble in the level
indicator. Refer to figure 2. The level indicator is located on the back of the balance. Make sure the fixed foot is pressed down
against the bench top and use the three adjustable feet to level the balance.

Figure 2 – Bottom and top views

Install the Weighing Pan:

Remove the protective plastic coating from the weighing pan and place the pan on top of the balance.

Connect to an Electrical Outlet:

The balance is supplied with either a 120V or 220V remote power supply. Before making any connections, verify that the power
(VAC) requirement shown on the remote power supply is compatible with the AC power outlet to which the balance will be
connected.

First plug the power supply’s connector into the balance’s rear panel receptacle. Then plug the power supply into the AC outlet. The
balance will automatically turn on and go through a self check sequence. During the self check the balance’s programming and
display segments are verified. When the self check is complete the balance will remain in a Stand By mode which is designated by a
single dot in the lower right corner of the display.

Figure 3 – Rear view

Turn the balance on:

Press the ON/OFF button one time to turn the balance on and observe the following turn on sequence.

Figure 4 – Turn on sequence


Allow the balance to warm up for at least 1 hour, then follow the calibration instructions to calibrate the balance.